22–26 Jun 2015
Europe/Vienna timezone

T3.1-P18. New Beta-Gamma coincidence system for measurement of radioxenon isotopes using well-type NaI(Tl) and plastic scintillator detectors

Not scheduled
Poster 3. Advances in sensors, networks and processing

Speaker

Hossein Afarideh (AMIRKABIR University of Technology)

Description

In order to detect radioxenon isotopes, a homemade geometry detection system based on beta-gamma coincidence has been developed. The system consists of a well-type NaI(Tl) detector that surround plastic scintillator in the solid angle near the 4π. The well-type NaI(Tl) detector measure gamma or X-ray and plastic scintillator detect electron particles. The beta detector was calibrated using 137Cs gamma Compton scattering in the plastic scintillator. It has been found that beta calibration curve depends to 137Cs position therefore; to suppress this effect an effective point for beta calibration was determined and verified with 131mxe conversion electron. The efficiency of gamma detector was measured using standard point sources. Since the detection system has a volume source therefore, the detection system was simulated by Gate7 and correction factor was obtained by least squares fitting of simulation results to experiment results. The accuracy of detection system performance was checked by injection of 214Pb (daughter of 226Ra) gas source.

Primary author

Hossein Afarideh (AMIRKABIR University of Technology)

Presentation materials

There are no materials yet.